%0 Journal Article %T Accuracy of minimally invasive percutaneous thoracolumbar pedicle screws using 2D fluoroscopy: a retrospective review through 3D CT analysis %A Winder, Mark J. %A Gilhooly, Paul M. %J Journal of Spine Surgery %D 2017 %B 2017 %9 %! Accuracy of minimally invasive percutaneous thoracolumbar pedicle screws using 2D fluoroscopy: a retrospective review through 3D CT analysis %K %X Background: In the literature, there is a large variation in the reported misplacement rates of pedicle screws. The use of minimally invasive surgical techniques is increasing and as such there has only been a small amount of data to look at the misplacement rate of percutaneously inserted thoracic and lumbar pedicle screws. Methods: A retrospective analysis of post-operative computed tomography (CT) scans were performed on 108 patients who underwent minimally invasive percutaneously inserted thoracic and lumbar pedicle screws by a single surgeon. Analysis of the screw trajectory using strict guidelines was performed using multiplanar reconstruction CT scan data to determine the accuracy of the pedicle screws. Results: A total of 614 screws were inserted in vertebral levels T2 to S1. Twenty-five (4.07%) screws were considered misplaced having breached the pedicular cortex. Thoracic pedicle screws had a statistically higher misplacement rate than lumbar pedicle screws (14.67% vs . 2.60% respectively, P Conclusions: Percutaneously inserted pedicle screws using 2D fluoroscopy offers a safe and accurate option for spinal stabilisation with an extremely low misplacement rate and morbidity. Overall, the low misplacement rates were equivalent and in most cases lower compared to open and computer assisted navigation techniques. However, we would recommend that given a misplacement rate of 14.67% for thoracic pedicle screws that computer assisted navigation may be able to offer further improvements in accuracy. %U https://jss.amegroups.org/article/view/3822 %V 3 %N 2 %P 193-203 %@ 2414-4630